Sponsor
This work was supported by the Environmental Protection Agency on Grant No. R811876.
Published In
Review of Scientific Instruments
Document Type
Article
Publication Date
1-1-1988
Subjects
Electric discharges -- Prevention, Mass spectrometers -- Design and construction
Abstract
An electrical discharge suppression system for a medium throughput (∼2 l/s) pumping line has been devised that works up to potentials of ±15 kV. This device permits atmospheric pressure ionization sources to be interfaced to high-resolution, magnetic sector mass spectrometers with source potentials of 6-10 kV
DOI
10.1063/1.1139855
Persistent Identifier
http://archives.pdx.edu/ds/psu/7637
Citation Details
Grange, A. H., O’Brien, R. J., & Barofsky, D. F. (1988). Discharge suppression system for a double focusing, atmospheric pressure ionization mass spectrometer. Review Of Scientific Instruments, 59(4), 656.